Developed Method for Low Concentration Determination of Uranium

Developed Method for Low Concentration Determination of Uranium

Authors

  • Ashraf El-Sayed, Mostafa M. Hamed

Keywords:

Uranium; Uv-Vis; Derivatives; Micells

Abstract

An improved spectrophotometric method of determination of U(VI) with 4-(2-pyridylazo) resorcinol (PAR) in a
mixed micellar medium (mixed cationic and nonionic surfactants) in presence of EDTA/NaF as masking agent
has been carried out in normal and derivative modes. The molar absorption coefficient (ε) and Sandell’s
sensitivity (S) of U(VI)-PAR complex at 541.0 nm are 4.33 x 104 mol-1 cm-1 , 0.0055 μg.cm-1
, respectively. The
proposed method has been successfully applied to the determination of uranium in tap water, river water,
seawater and synthetic sample.