Assay Determination of Tranexamic acid in Pharmaceutical Dosage Form (Tablet) Using HPLC and ELS Detector

Assay Determination of Tranexamic acid in Pharmaceutical Dosage Form (Tablet) Using HPLC and ELS Detector

Authors

  • Kiran R. Patil, Vipul P. Rane, Jaiprakash N. Sangshetti, Devanand B. Shinde

Keywords:

ELS detector; Tranexamic acid; Pharmaceutical preparations; Validation

Abstract

A simple, rapid, and precise method is developed for the assay determination of tranexamic acid (Tr)
in pharmaceutical dosage form. The method was developed with HPLC, ELS detector and Alltima
cyano 100A (250 x 4.6 mm, 5µ) column, the mobile phase composed of aqueous solution of 0.05%
v/v trifluoroacetic acid and acetonitrile (60:40 v/v, pH 2.85). The instrumental settings are flow rate of
0.5 mL min-1, column temperature at 300
C and detector drift tube temperature set at 970
C, a detector
nitrogen gas flow set at 2.6 SLPM, gain 1 and 5 µL sample injection volume. Theoretical plates were
20007. Tailing factor was 1.38. The described method shows excellent linearity over a range of 50 to
150% of assay concentration. The correlation coefficient was 0.9987. The relative standard deviation
of six measurements for peak area and retention time less than 2% and 0.5% respectively. The
proposed method was found to be suitable and accurate for assay determination of Tr in
pharmaceutical preparations.